Applicable changes of AS4775:2025
Key things unchanged:
- Critical measurements, flow rates and pressure remain unchanged – such as shower flow rate to be 75.7L/min at 210kPa; eye/face washer to be 11.4L/min at 210kPa for ≥15min.
Possible major effects:
- Clause 4.2 AS/NZS 4020 required for drinking-water-contact components
- Appendix A.6.1 requires conformity marking –
- Manufacturer’s trademark;
- Conformity mark;
- Licence no.
- Batch id;
- Specification no.
- Appendix A.6.2 requires Assembly markings to be durable or incorporated, including Inlet; Outlet; max/min pressures and temperatures; flow rates and model no.
- Clause 14 Detailed requirements for weekly testing to be maintained (weekly risk assessment)
Minor effects:
- Alarm system needs to be reviewed; new version requires the alarm can be central monitored, and sound to be distinguishable in tone and pattern to fire and evacuation
- Temperature minor change, from 15.6-37.8℃ -> 15.0-37.0℃
- Clause 16.1 mentions laminar flow should be used in microbiological laboratories or corrosive environments; otherwise, either aerated or laminar flow is acceptable for general circumstances.